Page 2
For the customers in Europe switch the power supply off, or disconnect the power The manufacturer of this product is Sony Corporation, plug. 1-7-1 Konan, Minato-ku, Tokyo, 108-0075 Japan. The Authorized Representative for EMC and product safety is Sony Deutschland GmbH, Hedelfinger Strasse 61, 70327 Stuttgart, Germany.
Page 3
1-2 and CISPR11, Class B, Group 1) Rear The equipotential terminal which brings the 6. Model LMD-2451MD is a monitor intended for use various parts of a system in a medical environment to display pictures from to the same potential.
Page 4
LMD-2451MD. Guidance and manufacturer’s declaration-electromagnetic emissions The LMD-2451MD is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-2451MD should assure that it is used in such an environment. Emission test Compliance...
Page 5
Guidance and manufacturer’s declaration - electromagnetic immunity The LMD-2451MD is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD- 2451MD should assure that it is used in such as environment. IEC 60601 test...
Page 6
Guidance and manufacturer’s declaration - electromagnetic immunity The LMD-2451MD is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD- 2451MD should assure that it is used in such as environment. IEC 60601 test...
Page 7
Recommended separation distances between portable and mobile RF communications equipment and the LMD- 2451MD The LMD-2451MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the LMD-2451MD can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (Transmitters) and the LMD-2451MD as recommended below, according to the maximum output power of the communications equipment.
Table of Contents Precaution .............. 9 On Safety ............9 On Installation ............ 9 About the LCD Display Panel ......9 About the Fluorescent Tube ....... 9 On Cleaning ............9 Disposal of the Unit ......... 10 Recommendation to Use more than One Unit .................
• To disconnect the AC power cord, pull it out by Sony dealer. grasping the plug. Never pull the cord itself. • The socket-outlet shall be installed near the equipment and shall be easily accessible.
If you have any questions about this unit, contact your Auto chroma/phase function authorized Sony dealer. The chroma and phase of the decoder are automatically adjusted with the auto chroma phase function. High image-quality/high-resolution WUXGA On Fan Error LCD panel An WUXGA high-resolution (1,920 ×...
Page 11
The number of the DVI input connectors can be Aspect setting increased by installing the optional input adaptor into You can set the monitor to 4:3 or 16:9 display mode the optional input slot. according to the input signal. To view more than SXGA signals when the DVI input is Scan function selected, use the cable within 3 m (118 inches) in...
Location and Function of Parts and Controls Front Panel COMPOSITE COMPONENT – – – HD15 – – USER MEM a Tally lamp Adjusts the color intensity. You can check the status of the monitor by the color of Press the + button to increase the color intensity or the – the tally lamp.
j Input select buttons Press the button to monitor the signal input to each connector. A-1, A-2, B-1 and B-2 buttons are used when an optional input adaptor has been installed in the option slot. COMPOSITE button: to monitor the signal through the COMPOSITE IN connector Y/C button: to monitor the signal through the Y/C IN connector...
Rear/Bottom Panel 3 4 5 6 7 qd qf /I (Equipotential/Function Earth) terminal To use the external sync signal, press the function button (equipotential) terminal that EXT SYNC is assigned (F1 button at the factory Connects the equipotential plug. setting). I (function earth) terminal Connects the earth cable.
Page 16
o DVI-D input connector (DVI-D) (power) switch The power is turned on or off. Inputs DVI Rev.1.0 applicable digital RGB signal. The monitor is turned on by pressing side To view the signals of the SXGA and higher resolution when the DVI input is selected, use the cable within 3 m l PARALLEL REMOTE connector (modular (118 inches) in length.
Insert the DC IN connector into the DC 5V/24V IN Connecting the AC connector on the bottom of this unit until it locks. Power Cord Connect the supplied AC power cord as illustrated. Two kinds of AC plug holders are supplied with this unit.
Installing the Input Removing the Adaptor Connector Cover Before installing the input adaptor, disconnect the power When the unit is shipped from the factory, a connector cord. cover is attached to the PARALLEL REMOTE connector and the SERIAL REMOTE connector (RJ- Remove the panel of the optional input slot.
Selecting the Default Settings When you turn on the unit for the first time after – purchasing it, select the area where you intend to use this – unit from among the options. – The default setting values for each area –...
2 If LATIN AMERICA is selected: Selecting the Menu PAL&PAL-N area Language L A T I N A M E R I C A P A L & P A L - N A R E A Argentina A R G E N T I N A P A R A G U A Y Paraguay U R U G U A Y...
The setting items (icons) in the selected menu are Using the Menu displayed in yellow. USER CONFIG – SYSTEM SETTING The unit is equipped with an on-screen menu for making M AT R I X : C O M P O N E N T L E V E L : xxxx various adjustments and settings such as picture control, N T S C S E T U P :...
Page 22
The menu icon presently selected is shown in About the memory of the settings yellow and setting items are displayed. The settings are automatically stored in the monitor memory. USER CONFIG – SYSTEM SETTING M AT R I X : C O M P O N E N T L E V E L : xxxxx N T S C S E T U P :...
Loading USER MEMORY Adjustment Using the Menus You can load the picture settings saved in the USER MEMORY menu (on page 32). Items The screen menu of this monitor consists of the COMPOSITE following items. COMPONENT – STATUS (the items indicate the –...
MATRIX For the video input COMPONENT LEVEL NTSC SETUP STATUS 1/2 GAMMA F O R M AT xxxxxxxxx FORMAT DISPLAY xxxxxxxx LANGUAGE C O L O R T E M P POWER SAVING C O M P O N E N T L E V E L xxxxx I/P MODE N T S C S E T U P...
COLOR TEMP/SPACE menu For the video input The COLOR TEMP/SPACE menu is used for adjusting the picture white balance or color space. USER CONTROL 1/3 You need to use the measurement instrument to adjust AU T O C H R O M A / P H A S E the white balance.
Page 26
For the DVI/HD15 input Submenu Setting SUB CONTROL Adjusts finely the adjustment range * The 1/3 menu cannot be adjusted. of the button on the front panel for CONTRAST, BRIGHTNESS, CHROMA and PHASE. USER CONTROL 2/3 • CONTRAST: Adjusts the picture S U B C O N T R O L contrast.
USER CONFIG menu Submenu Setting The USER CONFIG menu is used for setting the PICTURE CONTROL Adjusts to monitor the picture more clearly. system, multi display, function button, computer detect • SIZE H: Adjusts the horizontal and option DVI. size of the picture. The higher the setting, the larger the horizontal size of the USER CONFIG...
MULTI DISPLAY SETTING Submenu Setting LANGUAGE Selects the menu or message language from among seven USER CONFIG MULTI DISPLAY SETTING languages. M U LT I D I S P L AY E N A B L E : • ENGLISH: English M U LT I D I S P L AY: •...
About the function assigned to the Submenu Setting function button SUB INPUT Sets the input signal of the sub display. SELECT You can select from among SCAN COMPOSITE, Y/C, RGB, COMPONENT, DVI, HD15, OPTION Press to change the scan size of the picture. Press to A-1, OPTION A-2, OPTION B-1, switch between NORMAL scan (0% scan), OVER scan OPTION B-2, VIDEO WAVE and OFF.
Scan mode image Note If the APA operation does not finish correctly depending on the input signal, adjust DOT PHASE (page 27). I/P MODE Press the button to set the delay by the picture processing to the minimum level when interlace signal is input.
REMOTE menu Submenu Setting COMPUTER The appropriate preset memory is set for REMOTE DETECT the signal from DVI and HD15 input PA R A L L E L R E M O T E : connector. Select “PRESET1” for the standard computer signal.
USER MEMORY menu Submenu Setting MONITOR Set the monitor setting. IP ADDRESS: Sets the IP U S E R M E M O RY 1 / 3 address. x0 1 U S E R 0 1 SUBNET MASK: Sets the ·...
Page 33
The USER MEMORY setting menu appears. Press the – button to select “NAME”, then press the ENTER button. U S E R M E M O R Y 1 / 3 The menu for setting the user name appears. 0 1 U S E R 0 1 xS AV E : ·...
Troubleshooting Specifications Picture performance This section may help you isolate the cause of a problem and as a result, eliminate the need to contact technical LCD panel a-Si TFT Active Matrix support. Pixel efficiency 99.99% • The display is colored in green or purple t Select Viewing angle (panel specification) the correct input by pressing RGB or COMPONENT 89°/89°/89°/89°...
Page 35
Note BKM-229X are installed) Always verify that the unit is operating properly before Operating conditions use. SONY WILL NOT BE LIABLE FOR DAMAGES OF ANY KIND INCLUDING, BUT Temperature NOT LIMITED TO, COMPENSATION OR 0 °C to 35 °C (32 °F to 95 °F)
Pin assignment Available signal formats The unit is applicable to the following signal formats. PARALLEL REMOTE connector For details on the input signal available for HD15, DVI Modular connector and BKM-256DD, see page 37. (8-pin) Composite BKM- BKM- BKM- Pin number Functions System Component...
Page 37
Available HD15/DVI/BKM-256DD input signal format About the preset signal This unit has a preset memory for signals connected to the HD15 and DVI input connectors, and BKM-256DD. When a preset signal is input, the unit automatically detects the signal type and recalls the data for the signal from the preset memory to adjust it to an optimum picture.
Side Dimensions When an optional stand SU-560 is attached Front 110 (4 When an optional stand SU-560 is attached 602.4 (23 COMPOSITE CONTROL COMPONENT CONTRAST – PHASE – CHROMA – HD15 BRIGHT – MENU – ENTER 302 (12) USER MEM RETURN Mass Approx.